Pro X & USB Controller

Learn the benefits and advantages of the Aerogen Pro X and USB controller

Click on the blue markers

Click on image to show labels

Aerogen Pro-X Controller:

Portable power source for the Aerogen Solo with 30 minute and continuous modes

Aerogen Solo
Warning Indicator
Press and hold on/off for 1 SECOND
Power
Continuous ModeContinuous Mode
Press and hold on/off for 3 SECONDS
30 Minute Mode 30 Minute Mode
Press and hold on/off for 1 SECOND
Aerogen Solo
Quick-Set-Up
Easy-to-Use
Reusable
battery_icon
Powers
Multi-functional
Easily-Attached

*1 Second = 30 Minute Mode, 3 Seconds = Continuous Mode

Aerogen USB Controller:

Portable power source operated from USB ports on ventilators or any other medical equipment (not available in the US) or using the Aerogen AC/DC adapter.

The Aerogen USB controller has a choice of 30-minute and 6-hour operation modes. Simple to set up, the Aerogen USB Controller can be used multiple times, on different patients, throughout the hospital.3 A medical grade battery bank ⓘ is also available to buy as an option, which supplies 5 - 8 hours of operation.

Aerogen Solo
30 Minute Mode
Press and hold on/off for 1 SECOND
6 Hour Mode
Press and hold on/off for 3 SECONDS
Warning Indicator
Press and hold on/off for 1 SECOND
Power
Aerogen Solo
Powers
USB-Powered
man wearing a face mask
man wearing a face mask
man wearing a face mask
man wearing a face mask

*Press and hold for 1 second for 30 minute mode, or 3 seconds for continuous mode

** Powered From USB Ports on Ventilators, or any other medical electrical equipment, or plug sockets with an Aerogen AC/DC adapter

Troubleshooting

  1. Follow the controller AC/DC power cable from the controller and check that it is plugged in.
  2. Follow the Aerogen Solo cable from the controller to the Solo device and inspect for damage.
  3. Wiggle the cable at the device and controller attachments, watching for any warning lights.
    If any of the cables are damaged, replace them.
  4. Ensure proper orientation of the Aerogen Solo within the application.
    (Please refer to the Aerogen Solo instruction manual for approved Aerogen setups.)
  5. To clean the Aerogen Pro C controller, wipe it clean with an alcohol-based disinfectant.
    (Use alcohol-based disinfectant or quaternary ammonium compound-based disinfectant wipes. Please refer to Aerogen controller instructions for additional details.)